Logo Phone

UD-T3040T2(Receiver for UVR-T2)

UD-T3040T2(Receiver for UVR-T2)

Characteristic

Due to the distribution response characteristics in a wide range of flat ultraviolet wavelength regions, it is possible to measure the light intensity within the sensitivity range of 290-430nm without depending on the emission wavelength distribution of UV-LED, so the variation in emission wavelength of UV-LED Reduced measurement is possible.
The heat resistance has been increased from 40 ℃ to 60 ℃, which is currently on the market, making it the most suitable specification for lamp intensity management of conveyor type irradiation equipment used at high temperatures. Please use in combination with UVR-T2.

  • ・Measurements with reduced variation in UV-LED emission distribution are possible.
  • ・Spectral response characteristics in a wide range of flat ultraviolet wavelengths of 290-430nm.
  • ・Supports measurement of radiation limit(mW / ㎠), peak irradiance (mW / ㎠), and integrated value (mJ / ㎠).
PDFcatalog download

specification

specificationSpecification value
UD-T3040T2
Light receiving elementSilicon photodiode
Light receiving window diameterφ3mm
Measurement wavelength range300 ~ 390nm
Peak sensitivity wavelength約355nm
Illuminance measurement range
(mW/cm2)
Range-1:0.01~60.0
Range-2:0.1~600.0
Range-3:1.0~6000.0
Integrated light intensity measurement range
(mJ/cm2)
Range-1: 0.01~999.99
Range-2: 0.1~9999.9
Range-3 :1.0~99999.0
Analog output
1mV irradiance
(mW/ cm2)
Range-1: 0.03
Range-2: 0.3
Range-3: 3.0
data
sampling
Sampling time 50Hz: 10ms, 60Hz: 8.33ms
Sampling rate 50Hz: 50 times / second 60Hz: 60 times / second
Oblique light characteristics30 °: Within ± 5%, 60 °: Within ± 25%
Calibration accuracy(*1)Within ± 2%
Linearity(*2)Within ± 3%
Analog output voltage0~2Vmax. Common to each range
interfaceUSB2.0 (USB A connector-USB micro B connector)
Power supplyAAA battery x 3 (commercially available) / USB bus power
terms of useTemperature: 10-60 ℃ / Humidity: 85% R.H. or less (no condensation)
External dimensions
(W×D×H) (*3)
75×201×16mm
mass(*3)(*4)320g
Optional accessoriesExtension unit ZV-48 (2m)

(*1) For the value of the calibration standard with our calibration light source
(*2) When a light amount of 5% or more of full scale is incident in each measurement range, when zero correction is used
(*3) Without heat protection cover
(*4) Including battery